Cannula Zero
Instructions for sterilization of the cannula
Before washing, in the “iron washer”, it is necessary to separate the selector from the body of the cannula, in order to allow the removal of any pathogenic microorganisms inside the moving parts.
After washing, put the two pieces in a bag and place them into an autoclave machine at 121°C for 20 minutes.
It is important that the two pieces of the cannula are sterilized separately to avoid possible welding together.
The mirror must absolutely remain in its place to prevent the support from deforming and therefore making it difficult to re-insert it after sterilization.
Inserting the mirror
Inserting the mirror into the support, when you decide to replace it, is easier if done when hot (for example after an autoclave cycle) or in any case, if cold, with extreme delicacy.